EasyMax Voice
Self-Monitoring Blood Glucose System
Quick Reference Guide
Vue de la page 0
1 2 3 4 5 6 ... 16 17

Résumé du contenu

Page 1 - Quick Reference Guide

1 EasyMax Voice Self-Monitoring Blood Glucose System Quick Reference Guide

Page 2 - Installing Batteries

10 2. Put a test strip into the meter in the direction of the arrow. The meter turns on automatically. The icons and show themselves. 3. When t

Page 3

11 5. Gently squeeze and/or massage your fingertip until a round drop of blood on your fingertip. 6. Touch the blood drop at the tip of the t

Page 4

12 1. Massage the puncture area of forearm for a few seconds. 2. Press and hold the device with clear adjustable tip against the forearm. 3. Press the

Page 5

13 Discarding Used Lancets 2. Please discard used disposable lancets according to the recommendations of your healthcare professional. 1. Uns

Page 6

14 Viewing & Deleting Test Results The meter provides 7, 14, 30, 60 and 90 days averaging to help track your blood glucose trend soundlessly.

Page 7 - 1. Unscrew the Cap

15 Screen Messages And Troubleshooting Never make treatment decisions based on an error message. If you have any concerns, please contact your local d

Page 8 - 3. Recap the front cap

16 The ”Ht” and thermometer icon appears.Temperature is too high, outside the required range of 10°C - 40°C (50°F - 104°F). This alerts u

Page 9

17 Manufacturer: EPS Bio Technology Corp. 2F, No.49-2, Lane 2, Guang-Fu Rd., Sec. 2, Hsinchu City, Taiwan Tel: 886-3-5752522 Fax: 886-3-5752552 E-ma

Page 10

2 Installing Batteries 1. Open the battery door on the back of the meter by pushing the tab in the direction of the arrow and pulling t

Page 11

3 Setting The Time and Date—First Time Use Setting the current time and date in your meter is important if you intend to use the meter memory. 1

Page 12

4 Setting The Language and Volume—First Time Use 1. Press  to adjust the language.(En= English, Es = Spanish) 2. And press  to confirm the s

Page 13 - Discarding Used Lancets

5 Running A Control Solution Test You need the meter, a test strip, and control solution. 1. Put a test strip into the meter in the direct

Page 14

6 6. Touch the drop at the tip of the test strip. 7. The meter starts to count down from 5 seconds and then shows the result. 8. Do not r

Page 15

7 Inserting A Lancet Into The Lancing Device You must first load the lancet into the lancing device to get it ready for use. 1. Unscrew the

Page 16 - 10°C - 40°C (50°F

8 3. Recap the front cap.4. Select the desired penetration depth. Select 1-2 for soft or thin skin. 3 for average. 4-5 for thick or callouse

Page 17

9 5. Pull on the sliding barrel of the lancing device until it clicks and then release. Now the lancing device is ready. Do not pri