Checking Your Meter
A new check strip is included with your meter. The check strip tests only the function of the meter and
not the test strip’s ability to render a result (which would need a control solution), or the accuracy of
the system as a whole.
1. Make sure your check strip is
inserted. The meter will turn
on automatically.
2. The screen displays "OK".
3. Remove the check strip and
the meter will turn off
automatically.
Store the check strip in a safe
place.
Note:
When the display shows “FAL”, re-insert the check strip. If FAL
appears again, please call our Toll-Free number 1-866-203-2761.